Home ProductsDrug of Abuse Tests

One Step Marijuana Test Device Package Insert

One Step Marijuana Test Device Package Insert

One Step  Marijuana Test Device Package Insert
One Step  Marijuana Test Device Package Insert

Large Image :  One Step Marijuana Test Device Package Insert Get Best Price

Product Details:
Place of Origin: CHINA
Brand Name: SPAN
Certification: ISO CE
Model Number: MAR
Payment & Shipping Terms:
Minimum Order Quantity: 100 tests
Price: Negotiable
Packaging Details: 25 tests/box
Delivery Time: within 3-8working days(depends on your quantity)
Payment Terms: T/T, Western Union, MoneyGram
Supply Ability: 1000000 pcs per month
Detailed Product Description
Package: 20 Tests/box Specificity: 100%
Sensitivity: 100%

One Step Marijuana Test Device Package Insert
A rapid, one step test for the qualitative detection of THC metabolites in human urine.
For healthcare professionals including professionals at point of care sites
For in vitro diagnostic use only.

 

INTENDED USE
The THC One Step Marijuana Test Device is a rapid chromatographic immunoassay for the detection of 11-nor-9 -THC-9 COOH (THC metabolite) in human urine at a cut-off concentration of 50 ng/mL. This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. SUMMARY THC (9-tetrahydrocannabinol) is the primary active ingredient in cannabinoids (marijuana). When smoked or orally administered, it produces euphoric effects. Users have impaired short-term memory and slowed learning. Users may also experience transient episodes of confusion and anxiety. Long term relatively heavy use may be associated with behavioral disorders. The peak effect of smoking marijuana occurs in 20-30 minutes and the duration is 90-120 minutes after one cigarette. Elevated levels of urinary metabolites are found within hours of exposure and remain detectable for 3-10 days after smoking. The main metabolite excreted in the urine is 11-nor-9-tetrahydrocannabinol-9-carboxylic acid (9-THC-COOH). The THC One Step Marijuana Test Device is a rapid urine screening test that can be performed without the use of an instrument. The test utilizes a monoclonal antibody to selectively detect elevated levels of marijuana in urine. The THC One Step Marijuana Test Device yields a positive result when the concentration of marijuana in urine exceeds 50 ng/mL. This is the suggested screening cut-off for positive specimens set by the Substance Abuse and Mental Health Services Administration (SAMHSA, USA). PRINCIPLE The THC One Step Marijuana Test Device is a rapid chromatographic immunoassay based on the principle of competitive binding. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody. During testing, a urine specimen migrates upward by capillary action. Marijuana, if present in the urine specimen below 50 ng/mL, will not saturate the binding sites of the antibody coated particles in the test device. The antibody coated particles will then be captured by immobilized marijuana conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the marijuana level is above 50 ng/mL because it will saturate all the binding sites of anti-marijuana antibodies. A drug-positive urine specimen will not generate a colored line in the test line
region because of drug competition, while a drug-negative urine specimen will generate a line in the test line region because of the absence of drug competition. To serve as a procedural control, a colored line will always appear at the control line region if the test has been performed properly. REAGENTS The test device contains mouse monoclonal anti-Marijuana antibody-coupled particles and Marijuana-protein conjugate. A goat antibody is employed in the control line system. PRECAUTIONS  For healthcare professionals including professionals at point of care sites.  For in vitro diagnostic use only. Do not use after the expiration date.  The test device should remain in the sealed pouch until use.  All specimens should be considered potentially hazardous and handled in the same manner as an infectious agent.  The test strip should be discarded according to federal, state and local regulations. STORAGE AND STABILITY Store as packaged in the sealed pouch at 2-30°C. The test device is stable through the expiration date printed on the sealed pouch. The test device must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date. SPECIMEN COLLECTION AND PREPARATION
Urine Assay The urine specimen must be collected in a clean and dry container. Urine collected at any time of the day may be used. Urine specimens exhibiting visible precipitates should be centrifuged, filtered, or allowed to settle to obtain a clear specimen for testing.
Specimen Storage Urine specimens may be stored at 2-8°C for up to 48 hours prior to testing. For prolonged storage, specimens may be frozen and stored below -20°C. Frozen specimens should be thawed and mixed before testing.
MATERIALS
Materials Provided  Test devices  Disposable specimen droppers  Package insert
Materials Required But Not Provided  Specimen collection container  Timer  External controls DIRECTIONS FOR USE Allow the test device, urine specimen, and/or controls to equilibrate to room temperature (15-30C) prior to testing. 1. Bring the pouch to room temperature before opening it. Remove the test device from the sealed pouch and use it as soon as possible. 2. Place the test device on a clean and level surface. Hold the dropper vertically and transfer 3 full drops of urine (approx. 100l) to the specimen well (S) of the test device, and then start the timer. Avoid trapping air bubbles in the specimen well (S). See the illustration above. 3. Wait for the red line(s) to appear. The result should be read at 5 minutes. It is important that the background is clear before the result is read. Do not interpret the result after 10 minutes.



INTERPRETATION OF RESULTS (Please refer to the illustration above)
NEGATIVE:* Two lines appear. One red line should be in the control region (C), and another apparent red or pink line should be in the test region (T). This negative result indicates that the marijuana concentration is below the detectable level of 50 ng/mL. * NOTE: The shade of red in the test region (T) may vary, but it should be considered negative whenever there is even a faint pink line.
POSITIVE: One red line appears in the control region (C). No line appears in the test region (T). This positive result indicates that the marijuana concentration is above the detectable level of 50 ng/mL.
INVALID: Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test with a new test device. If the problem persists, discontinue using the test kit immediately and contact your local distributor. QUALITY CONTROL A procedural control is included in the test. A red line appearing in the control region (C) is the internal procedural control. It confirms sufficient specimen volume, adequate membrane wicking and correct procedural technique. A clear background is also required. Control standards are not supplied with this kit; however it is recommended that positive and negative controls be tested as good laboratory testing practice to confirm the test procedure and to verify proper test performance. LIMITATIONS 1. The THC One Step Marijuana Test Device provides only a qualitative, preliminary analytical result. A secondary analytical method must be used to obtain a confirmed result. Gas chromatography/mass spectrometry (GC/MS) are the preferred confirmatory methods.1,2 2. It is possible that technical or procedural errors, as well as other interfering substances in the urine specimen may cause erroneous results. 3. Adulterants, such as bleach and/or alum, in urine specimens may produce erroneous results regardless of the analytical method used. If adulteration is suspected, the test should be repeated with another urine specimen. 4. A Positive Result does not indicate level or intoxication, administration route or concentration in urine. 5. A Negative Result may not necessarily indicate drug-free urine. Negative results can be obtained when drug is present but below the cutoff level of the
test. 6. Test does not distinguish between drugs of abuse and certain medications.


PERFORMANCE CHARACTERISTICS
Accuracy A three way side-by-side comparison was conducted using the THC One Step Marijuana Test Device and a leading commercially available THC rapid test. Testing was performed on specimens previously collected from subjects presenting for Drug Screen Testing. Presumptive positive results were confirmed by GC/MS. The following results were tabulated: Method Other THC Rapid Test Total Results THC One Step Test Device Results Positive Negative Positive 143 0 143 Negative 0 157 157 Total Results 143 157 300 Total Agreement: 100%
When compared to GC/MS at 50 ng/mL, the following results were tabulated: Method GC/MS Total Results THC One Step Test Device Results Positive Negative Positive 119 24 143 Negative 3 154 157 Total Results 122 178 300 Relative Sensitivity: 98% Relative Specificity: 87% Accuracy: 91%
When compared to GC/MS at 25 ng/mL, the following results were tabulated: Method GC/MS Total Results THC One Step Test Device Results Positive Negative Positive 137 6 143 Negative 4 153 157 Total Results 141 159 300 Relative Sensitivity: 97% Relative Specificity: 96% Accuracy: 97%
Point of Care Accuracy A study was conducted using the same clinical specimens with ten percent (10%) distribution at 25% above and below the 50ng/mL cut-off at three geographically distinct point of care sites to determine the accuracy of the One Step Marijuana Test Device in the hands of point of care user. Forty (40) positive specimens and forty (40) negative specimens were tested on three (3) different lots of each product. The difference in sensitivity and specificity results obtained by the laboratory professional for the same clinical specimens compared to the results obtained by the point of care (untrained) user was insignificant. At a ninety-five percent (95%) confidence interval, the odds ratio for the point of care user versus the laboratory professional was 1 to 1 for sensitivity and specificity.
Analytical Sensitivity A drug-free urine pool was spiked with 11-nor-9-Tetrahytrocannabinol-9carboxylic acid at the following concentrations: 75 ng/ml, 62.5 ng/ml, 37.5 ng/ml, 25 ng/ml, and 0 ng/ml. The result demonstrates 100% accuracy at 50% above and 50% below the cut-off concentration. The data are summarized below: THC Concentration (ng/mL) Percent Of Cutoff n Visual Result Negative Positive 0 0 30 30 0 25 50% 30 30 0 37.5 75% 30 10 20 50 Cutoff 30 4 26
62.5 125% 30 3 27 75 150% 30 0 30


Specificity The following table lists compounds and their respective concentrations in urine that yield a positive result in the THC One Step Marijuana Test Device at 5 minutes.
Compound Concentration (ng/mL) Cannabinol 20,000 11-nor- 8 -THC-9 COOH 30 11-nor- 9 -THC-9 COOH 50  8 —THC 15,000  9 -THC 15,000
Precision A study was conducted at 3 physician’s offices by untrained operators using 3 different lots of product to demonstrate the within run, between run and between operator precision. An identical panel of coded specimens containing, according to GC/MS, no THC, 25% THC above and below the cut-off and 50% THC above and below the 50 ng/mL cut-off was provided to each site. For the specimens below the -25 % cut-off concentration, the 3 sites demonstrated 98% agreement with each other. For the –25% to +25% cut-off specimens, the 3 sites demonstrated 83% agreement with each other. For specimens above the +25% cut-off concentration, the 3 sites demonstrated 100% agreement with each other. For all results, the 3 sites were found to have a 92% agreement with each other. Effect of Urinary Specific Gravity Twenty-six (26) urine samples of normal, high, and low specific gravity ranges were spiked with 25 ng/ml and 75 ng/ml of 11-nor-9-Tetrahytrocannabinol-9-carboxylic acid, respectively. The One Step Marijuana Test Device was tested in duplicate using the twenty-six neat and spiked urine samples. The results demonstrate that varying ranges of urinary specific gravity does not affect the test results.
Effect of the Urinary pH The pH of an aliquoted negative urine pool was adjusted to a pH range of 5 to 9 in 1 pH unit increments and spiked with 11-nor-9-Tetrahytrocannabinol-9-carboxylic to 25 ng/ml and 75 ng/ml. The spiked, pH-adjusted urine was tested with the THC One Step Marijuana Test Device in duplicate and interpreted according to the package insert. The results demonstrate that varying ranges of pH does not interfere with the performance of the test.
Cross-Reactivity A study was conducted to determine the cross-reactivity of the test with compounds in urine not associated with THC. The following compounds show no crossreactivity when tested with the THC One Step Marijuana Test Device at a concentration of 100 g/mL:
Non Cross-Reacting Compounds 4-Acetamidophenol Fenoprofen Pentazocine Acetophenetidin Furosemide Pentobarbital N-Acetylprocainamide Gentisic acid Perphenazine Acetylsalicylic acid Hemoglobin Phencyclidine Aminopyrine Hydralazine Phenelzine Amitryptyline Hydrochlorothiazide Phenobarbital Amobarbital Hydrocodone Phentermine Amoxicillin Hydrocortisone L-Phenylephrine Ampicillin O-Hydroxyhippuric acid -Phenylethlamine Ascorbic acid 3-Hydroxytyramine -Phenyllethylamine D,L-Amphetamine Ibuprofen Phenylpropanolamine
L-Amphetamine Imipramine Prednisolone Apomorphine Iproniazid Prednisone Aspartame (-) Isoproterenol Procaine Atropine Isoxsuprine Promazine Benzilic acid Ketamine Promethazine Benzoic acid Ketoprofen D,L-Propanolol Benzoylecgonine Labetalol D-Propoxyphene Benzphetamine Levorphanol D-Pseudoephedrine Bilirubin Loperamide Quinidine Brompheniramine Maprotiline Quinine Caffeine Meprobamate Ranitidine Cannabidiol Methadone Salicylic acid Chloralhydrate Methoxyphenamine Secobarbital Chloramphenicol (+) 3,4-Methylenedioxyamphetamine Serotonin (5Hydroxytyramine) Chlordiazepoxide
Chlorothiazide (+) 3,4-Methylenedioxymethamphetamine Sulfamethazine () Chlorpheniramine Sulindac Chlorpromazine Methylphenidate Temazepam Chlorquine Methyprylon Tetracycline Cholesterol Morphine-3--Dglucuronide Tetrahydrocortisone, 3 Acetate Clomipramine
Clonidine Nalidixic acid Tetrahydrocortisone 3 (-D glucuronide) Cocaine hydrochloride Nalorphine Codeine Naloxone Tetrahydrozoline Cortisone Naltrexone Thebaine (-) Cotinine Naproxen Thiamine Creatinine Niacinamide Thioridazine Deoxycorticosterone Nifedipine D, L-Thyroxine Dextromethorphan Norcodein Tolbutamine Diazepam Norethindrone Triamterene Diclofenac D-Norpropoxyphene Trifluoperazine Diflunisal Noscapine Trimethoprim Digoxin D,L-Octopamine Trimipramine Diphenhydramine Oxalic acid Tryptamine Doxylamine Oxazepam D, L-Tryptophan Ecgonine hydrochloride Oxolinic acid Tyramine Ecgonine methylester Oxycodone PrD, L-Tyrosine (-) Y Ephedrine Oxymetazoline Uric acid Erythromycin p-Hydroxymethamphetamine Verapamil -Estradiol Zomepirac Estrone-3-sulfate Papaverine Ethyl-p-aminobenzoate Penicillin-G
BIBLIOGRAPHY
1. Hawks RL, CN Chiang. Urine Testing for Drugs of Abuse. National Institute for Drug Abuse (NIDA), Research Monograph 73, 1986 2. Baselt RC. Disposition of Toxic Drugs and Chemicals in Man. 2nd Ed. Biomedical Publ., Davis, CA. 1982; 488



Printed in China DN: R015454-04 Eff. Date: 2002-12-18

 

 

Span Biotech Ltd. is a research based company for rapid tests, with strong support from National Key Laboratory of Technology Projects of 10th and 11th Five-Year Plan and Faculty of Life Sciences of HuBei University. SpanBio also housed a R&D team that is developing gene recombination, cell cultivation and protein purification techniques. SpanBio pays strict attention on rapid tests for human being, animal diseases and food safety detection. It provides a number of customized services to professional distributors and partnering affiliates with excellent quality, competitive prices and super service.

 

Our mission:

 

  • Always best of all and always pay attention to innovation.
  • Special customized service tightly following customers’ requests.
  • Integrated excellent quality, competitive prices and super service together.

 

 

 

Rebecca Yan
 
Int'l Business Manager
Span Biotech Ltd.
Tel: +86(755)89589611
WhatsAPP:+8618823462100 (Wechat)
Web:www.spanbio.com

 

 

Contact Details
SPAN BIOTECH LTD.

Contact Person: Ms. Anna Lee

Tel: +86-755-89589611

Fax: 86-755-89580096

Send your inquiry directly to us
www.spanbio.com